Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 23
Heart Rate Display Test
Tools Needed ECG simulator
ECG Cable (3-lead, 5-lead or 12-lead)
Test Setup Press the power switch to turn on the Propaq MD. Press Lead quick access key until II displays.
Connect the ECG leads to the ECG simulator.
Connect the ECG cable to the unit.
Do this... Observe this... Pass/Fail
1. Set the ECG Simulator to 120BPM. The Heart Rate displays as 120 +/- 4 bpm
oo