CHAPTER 1MAINTENANCE TESTS
26 9650-0801-01 Rev. A
Leakage Current Test
Tools Needed See the manufacturers instructions or supplied specifications for the leakage tester you use.
Setup See the manufacturers instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.
Maximum Leakage Acceptance Limits
Patient Leakage Mains on Applied Part
Normal Single Fault Single Fault
ECG 10
50 100
MFC 10
50 100
Normal Single Fault
Earth Leakage 2500
5000