Removing the EtCO2 and NIBP Modules
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 61
9 Remove five (5) screws and washers from
the NIBP unit to release it.
Note: During reassembly, torque screws to
6 in-lbs.
10 Remove multi-wire EtCO
2
cable with
ferrite.
11 Disconnect the brown SpO
2
flex cable as
shown.
12 Disconnect the NIBP tubing from the side
panel in the area illustrated.
13 Remove the NIBP module from the unit.