Dynatech MP-7000 Manual de servicio Pagina 74

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 92
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 73
CHAPTER 4LIFE CYCLE COUNTERS
70 9650-0801-01 Rev. A
CO2
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.
SpO2
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>SpO2>Error Counts.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.
Defibrillator/Pacer
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>Defib/Pacer>Error
Counts.
3. Select Reset CRC Counter. Select Ye s when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.
Vista de pagina 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 91 92

Comentarios a estos manuales

Sin comentarios