64 9650-0801-01 Rev. A
3 Lift the monitor board to expose the
multi-wire LCD cable on the back side of
monitor board and disconnect it.
4 Remove three (3) screws and washers
from the NIBP unit.
Note: During reassembly, torque screws to
6 in-lbs.
5 Lift the SpO2 unit straight up to disconnect
it from the monitor board and remove it
from the unit.
6 Lift out the monitor board.