Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 25
10. On the Propaq MD, press the Alarm
Suspend key ( ).
Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.
oo
10. Set simulator to 160 BPM or higher.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
HIGH ALARM message displays.
oo
11.
On the Propaq MD, press
the
Alarm Silence/Reset button
()
.
Alarm is suspended for 90 seconds.
The alarm suspension timer appears
in black.
oo
Do this... Observe this... Pass/Fail