Disassembling the Main Unit
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 53
6 Disconnect the battery communication
cable (yellow, orange, brown, red)
connector from the monitor board.
7 Disconnect the monitor/defibrillator flex
cable from the monitor board.
Note: Be careful to pull the connector
straight out. Rocking the connector or
disconnecting it at an angle may result in
bent pins.
8 Disconnect the three power cable
connectors (battery (black), AUX (white),
and defibrillator (red)).
9 Press the tabs on either side of the two
printer cables to disconnect them.
10 Use an orange stick to release the tabs on
either side of silver ZIF connector between
the dock connector and the monitor board.
Note: Release one side at a time.
battery
cable
flex cable
p
cables
three power
cables
three power
cables
p
three power
cables
three power
cables
printer cables
NIBP
hoses